CS BOX HFA CS Analyzer
Carbon And Sulphur Analyzer for Inorganic Samples by High Frequency Induction Furnace
UOP703-09 DIN 24935 DIN EN ISO 9556 ISO 13902 ISO 4935
- Iron
- Steel
- Cast Iron
- Refractories
- Rocks
- Cements
- Catalysts
Carbon And Sulphur Analysis by High Temperature Tube Combustion Technique
- User friendly design of Sylab, where furnace is isolated from
the infrared analyzer to improve the life of the NDIR analyzer - Automatic crucible loading with tool free auto cleaning of the
combustion tube - Fully Computer controlled operation through software
- Two individual solid state IR detectors for Carbon and sulphur
- Powerful Induction furnace for Inorganic samples analysis
- Single and Multi-point calibration facility
- Vertical furnace design with single sample auto loading
Item: CSBox-HFA
Application: CSBox-HFA is equipment for the carbon and sulfur determination in inorganic and metals
Measuring Technique: High Quality infrared Cells for precise and stable signals and performances in Analysis of Iron, Steels, Cast iron, Refractories , Rocks, Cement, Catalyst Etc.,
Complies:
Measuring Range:
Practical Analysis Range for a 1g Sample:
C: 0.0006 to 4%
S:0.0001 to 1%
Practical Analysis Range for a 50mg Sample:
C: 0 to 100%
S: 0 to 20%
Accuracy: ± 1% relative (Most Samples)
Sensitivity:0.00001% (0.1 ppm)
Analysis time: 20 – 60Sec typical
Sample Weight: 100-200mg
Furnace:
Furnace type: New Generation of High Frequency Solid State Furnace
HF Furnace: High Power,3.3kW New Generation HF Furnace 100% discreet components with automatic combustion tube cleaning.
Frequency: 1MHz
HF Power: The HF furnace controlled by 1% steps from 0 to 100%,this feature is not required for C/S determination in metals but interesting for some CO2 and SO2 determinations in inorganic under neutral atmospheres
Cleaning: The CSBox-HFA has an automatic combustion Tube cleaning, and no tools are required for combustion tube exchange (glove-box technology)
Detection System:
Detection Method: Non-Dispersive ,100% Static infrared absorption
Detector type:
General Requirements:
Ambient Conditions: 15 to 35 deg C
20 – 80% Humidity (not condensing)
Power Supply: 230V ˜(+10/-15% at Max load);50/60 Hz, Single Phase
Required Gas oxygen 99%
Required Accessories: Computer, Monitor, Balance